Cookies sur les sites Web Omega Nous utilisons des cookies sur ce site Web. Ces cookies sont indispensables au bon fonctionnement de notre site. Si vous continuez sans modifier vos paramètres, nous supposerons que vous êtes heureux de recevoir tous les cookies de notre site Web. Pour de plus amples informations à ce sujet, veuillez cliquer ici
Click on this link if you need more information about Pt100.
RTD 100 Ω sensor with built-in
transmitter is programmable by
a computer. The configuration
kit required for programming is
M12TX-CONFIG. This unique
probe is ideal for areas with space
limitations where traditional head
connections are too large to fit.
The M12 thread design offers a
secure industrial connection.
Specifications Body: Stainless steel AISI 316L Probe Length: 13 mm (0.51") and 24 mm (0.94") Diameter: Ø 3 mm (0.1") Probe Type: Thermowell AISI 316L SS Probe Minimum Bending Radius: Three-times the outer diameter [except the sensing tip which length is 30 mm (1.2")] Connection: 1/8 NPT male thread Sensor: RTD PT100 class A up to 300°C in accordance to IEC751 Range: -50 to 120°C (-58 to 248°F) Tmax. Electronic: 80°C (176°F) Sensor Break Monitoring: Selectable: Upscale (>21.0 mA) or downscale (<3.6 mA) action Sensor Short-Circuit: Fixed to downscale (<3.6 mA) action Output: Signal: 4 to 20 mA Permissible Load: 700 @ 24 Vdc [RLo= (Vsupply-8,5)/0,020] Default Factory Setting for 4 to 20mA output: 0 to 100°C Response Time (90%): <50 ms Isolation In-Out: Non-isolated Power Supply: 8.5 to 32 Vdc (polarity protected) Environment Conditions: Temperature: -40 to 80°C (-40 to 176°F) for plastic body Relative Humidity: 0 to 100% EMC: In accordance to EN 61326 Degree of Protection: IP65 and IP67 in accordance to IEC60529 Accuracy: Transmitter: Maximum of ±0.2°C or ±0.2% of span Sensor: Class A in accordance to IEC751 Temperature Influence [Deviation from 20°C (68°F)]: Maximum of ±0.3°C/25°C (77°F) or ±0.3% of span/25°C (77°F) Supply Voltage Influence: Negligible Range Configurations: It is possible set the input temperature range (span) by the M12TX-CONFIG configuration kit (PC with OS Windows required) Zero Adjustments: Any value between -50 to 50°C Minimum Span: 50°C (122°F) [if the zero value is set between one of these values: -40°C (-40°F), -20°C (-4°F), 0°C (32°F), 20°C (68°F), 40°C (104°F), the minimum span is 20°C (68°F) rather than 50°C (122°F)] Sensor Error Compensation: Over 2 points (maximum 1% of span) Factory Setting: 0 to 100°C/sensor break >21 mA (upscale) Response Time: Diameter 3 mm <3.5 seconds, diameter 6 mm <13 seconds (test in water to IEC751- time for reaching 63.2% of instantaneous temperature change)
† Tous les Montants indiqués Dans EUR I> Notez: Comes complete with operator’s manual.
Manuels de Produit:
Télécharger
M12 Series - Temperature Transmitter with 100 Ohm RTD
Télécharger
M12TX-CONFIG - Configuration Kit for M12TX