Cookies sur les sites Web Omega Nous utilisons des cookies sur ce site Web. Ces cookies sont indispensables au bon fonctionnement de notre site. Si vous continuez sans modifier vos paramètres, nous supposerons que vous êtes heureux de recevoir tous les cookies de notre site Web. Pour de plus amples informations à ce sujet, veuillez cliquer ici
Click on this link if you need more information about Data Acquisition .
The OM-WEB-TC is a thermocouple measurement device with built-in
web server. Simply connect the device to an Ethernet port or hub
and view current data using a standard web browser.
The OM-WEB-TC’s embedded web interface provides access to current
data and configuration settings using a standard web browser.
Browse to the device’s home page by entering the URL that is printed
on the device into the browser. View current thermocouple
measurements and channel data, and configure hardware options
from the device’s home page. Only one user can change configuration
options on the device at a time. The web interface is built into the
device’s firmware, and does not need to be installed on a computer.
No external software is required other than a web browser and a
TCP/IP connection. Connect the device’s 10Base-T Ethernet port to
a local or wide area network using the supplied ethernet cable, to a
single computer through a hub using the supplied ethernet cable, or
directly to a computer using a standard CAT-5 crossover cable.
The web browser used to access the OM-WEB-TC’s web interface
must support JavaScript.
The OM-WEB-TC provides 8 thermocouple channels and
8 digital I/O channels. The included external power supply (ac adaptor)
is used to provide power. On-board LEDs display the status of
communications and external power. All hardware configurable
options are selectable with the web browser or the included Instacal
utility software. Network configurable options are settable
using Instacal. When using Instacal, if the login settings have been
changed from the default, a login name and password are required to
change the configuration settings. The OM-WEB-TC provides
8 differential thermocouple input channels. A 24-bit analog-to-digital
(A/D) converter is provided for each pair of analog inputs. Thermocouple
types J, K, T, E, N, R, S, B are supported. The thermocouple type
is software programmable for each channel. Four cold junction
compensation (CJC) sensors are provided for thermocouple
measurements. Each CJC sensor is dedicated to two thermocouple input
channels. An open thermocouple detection feature lets you detect a
broken thermocouple. An on-board microprocessor automatically
linearizes the measurement data.
The OM-WEB-TC features 8 independent temperature alarms,
Each alarm controls an associated digital I/O channel as an alarm
output. The input to each alarm is one of the temperature input
channels, the output of each alarm is software configurable as an
active high or low. The user configurable threshold conditions
activate each alarm. When an alarm is activated, the associated
DIO channel is driven to the active output state selected.
Eight digital I/O channels are provided to communicate with
external devices and to generate alarms. The digital bits are software
programmable for input or output. The digital output voltage is switchselectable
for 3.3V or 5V logic. A screw terminal is provided for
pull-up or pull-down configuration. The OM-WEB-TC web interface
displays current data read from the device and does not log or store
historical data. Use the included TracerDAQ software to log or
graphically trend your data. The OM-WEB-TC module ships with
an impressive array of software, including the new TracerDAQ®, a
full-featured, out-of-the-box data logging, viewing, and analysis
application; Universal Library™; ULx for NI LabVIEW (comprehensive
library of VIs and example programs compatible with 32- and 64-bit
LabVIEW v8.5 through 2011) and InstaCal™ installation, calibration,
and test utility–powerful software solutions for programmers and
nonprogrammers alike. These modules operate under Microsoft
Windows XP/VISTA/7 (32-bit and 64-bit) operating systems.
Specifications
ANALOG INPUTS
A/D Converter: 4 dual 24-bit sigma delta A/D converters Input Isolation: 500 Vdc min between
field wiring and USB interface Number of Channels: 8 differential temperature inputs Differential Input Voltage Range: ±0.080V Absolute Maximum Input Voltage:
±25V (power on), ±40V (power off) Throughput Rate: 2 samples/sec
max for all active channels Input Impedance: 5 GΩ min Input Leakage Current:
105 nA max (with open thermocouple detection enabled) Normal Mode Rejection Ratio: 90 dB min Common Mode Rejection Ratio: 100 dB min Warm-Up Time: 30 minutes max Open Thermocouple Detection: 3 sec open detection time
(max) CJC Sensor Accuracy:
-0.75 to 0.5°C max (15 to 35°C);
-1.5 to 1.25°C max (0 to 55°C)
DIGITAL I/O
Number of Digital I/O Channels: 8 Type: CMOS Configuration: Each DIO bit can be independently configured for input
or output. Switch selectable output voltages of 5V or 3.3V. Power on
reset is input mode unless bit is configured for alarm Pull-Up/Pull-Down Configuration: All pins are connected to 47 Ω
resistors (user configurable for pull-up mode to 5V or pulldown
mode). Digital I/O Transfer Rate (Software Paced): Digital Input: 50 port reads or single bit reads per
second (typical) Digital Output: 100 port writes or single bit writes
per second (typical) Input High Voltage: 4.0V min, 5.5V absolute max (5V mode);
2.64V min, 5.5V absolute max (3.3V mode) Input Low Voltage: 1.0V max, -0.3V absolute min
(5V mode); 0.66V max, -0.3V absolute min (3.3V mode) Output High Voltage: 4.3V min
(5V mode), 2.7V min (3.3V mode); IOH = -2.5 mA Output Low Voltage: 0.6V max (IOL = 2.5 mA) Temperature Alarms: 8 (one per digital I/O line)
NETWORK
Ethernet Compliance
Device Type: IEEE 802.3 Ethernet 10Base-T Device Compatibility: IEEE 802.3-2003 10 Mpbs Media
Access Control Ethernet Connection Ethernet Type: 10Base-T Connector: RJ-45, 8 position Cable: CAT-5 shielded, unshielded twisted pair Length: 100 m (328') max MAC Address: 00:12:71:Cx:xx:xx where xxxxx is the device's
serial number Network Factory Default Settings IP Address: 192.168.0.101 Subnet Mask: 255.255.255.0 Gateway: 192.168.0.1 DHCP Setting: Enabled User Name: "webtc" Password: "omega" Web Server: Enabled Network Protocols Protocols Implemented: IP, ARP, ICMP, DHCP, UDP, TCP,
NBNS, HTTP UDP Messaging Protocol: UDP port 54211 TCP Downloading Protocol: TCP port 54267 HTTP 1.0 Alternate Port: TCP
port 49152-65535 (not including 54267) Network Name: "webtc_xxxxx", where xxxxx is
the device's serial number Max Number of Simultaneous HTTP Connections: 3 Max Number of Non-HTTP TCP Sockets: 5 Network Security
Security Implementation: IP address based session manager with username/password login for configuration and control
transactions (data is not secured) Session Timeout: 5 min with no activity Username/Password Encryption:
Base64 (the default web page does not support
encryption if Javascript is disabled in the web browser) Vulnerabilities: Denial of service attacks, username/password
spoofing, script probing and simple decryption
GENERAL
Memory: EEPROM (512 bytes for
sensor configuration); FLASH (2 MB
for device configuration and website
storage) Microcontroller: One high-performance 8-bit
RISC microcontroller and one high-performance 16-bit RISC microcontroller Power Supply Voltage (Supplied by Included
External Power Supply): 5 Vdc ±5% Power Supply Current (Supplied by Included
External Power Supply): 440 mA max User Output Voltage (5V):
4.65V min to 5.25V max Dimensions:
127 L x 89 W x 36 mm D
(5.0 x 3.5 x 1.4") Input Connections: Screw terminal blocks
(accept 16 to 30 AWG wire) Operating Temperature: 0 to 55°C
(32 to 131°F), 0 to 90% RH non-condensing Storage Temperature: -40 to 85°C (-40 to 185°F)
† Tous les Montants indiqués Dans EUR I> Notez: Comes complete with ethernet cable, 100 to 240 Vac AC adaptor with USA plug, quick start guide, software and operator’s manual on CD.
We noticed you are from the US. Visit your local site for regional offers and live support.